Cerumen Occlusion Lowers Infrared Tympanic Membrane Temperature Measurement

44Citations
Citations of this article
12Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Objective: To examine the effect that cerumen occlusion of the ear canal has on infrared tympanic membrane temperature measurement. Methods: A prospective, randomized, single‐blind human study was carried out in a university hospital observation unit. The subjects were a convenience sample of human volunteers aged 18 years or older who did not have cerumen occlusion or scarred tympanic membranes. A paraffin‐coated human cerumen plug was placed in one randomly chosen ear, and after 20 minutes of equilibration the temperature in each ear was measured with an infrared thermometer. Analysis of the difference in mean temperature between the occluded and nonoccluded ears was by Student's paired t‐test. Results: Infrared tympanic membrane temperatures were measured in 43 subjects aged 21 to 58 years. The mean temperature of the occluded ear canal was 0.3°C lower than that of the opposite ear canal (p = 0.0001, 95% CI 0.16–0.45°C). Conclusion: Cerumen occlusion of the ear canal causes underestimation of body temperature measured by infrared tympanic membrane thermometry. © 1995 Society for Academic Emergency Medicine

Cite

CITATION STYLE

APA

Doezema, D., Lunt, M., & Tandberg, D. (1995). Cerumen Occlusion Lowers Infrared Tympanic Membrane Temperature Measurement. Academic Emergency Medicine, 2(1), 17–19. https://doi.org/10.1111/j.1553-2712.1995.tb03072.x

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free