Recommendations of megavoltage computed tomography settings for the implementation of adaptive radiotherapy on helical tomotherapy units

10Citations
Citations of this article
8Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Megavoltage computed tomography (MVCT) image quality metrics were evaluated on an Accuray Radixact unit to recommend scan settings for the implementation of a consistent adaptive radiotherapy program. Megavoltage computed tomography image quality was evaluated and compared to a kilovoltage CT (kVCT) simulator using a commercial cone beam computed tomography image quality phantom. Megavoltage computed tomographies were acquired on the Accuray Radixact using fine, normal, and coarse pitches, with all available reconstruction slice thicknesses, each of which were reconstructed using standard and iterative reconstruction (IR). Image quality metrics (IQM) were evaluated using DoseLab: automatically and manually calculated spatial resolution, subject contrast, and contrast-to-noise ratio (CNR). Scanning time was 15.6 s/cm for fine, 8.1 s/cm for normal, and 5.6 s/cm for coarse pitch. Automatically evaluated spatial resolutions ranged from 0.39, 0.41, to 0.42 lp/mm for standard reconstruction and from 0.24, 0.21, to 0.18 lp/mm for soft-tissue IR, respectively, with general IR yielding values in between these. Spatial resolution for kVCT was measured to be at least 0.42 lp/mm. Contrast was consistent across MVCT settings with 8.1 ± 0.2%, while kVCT contrast was 10.27 ± 0.05%. CNR was calculated to be 3.3 ± 0.4 for standard reconstruction, 7.4 ± 0.4 for general IR, and 12.0 ± 1.9 for soft-tissue IR. It was found that increasing reconstruction slice thickness for a given pitch does not improve IQMs. Based on the consistency of contrast metrics across pitch values and the only slightly reduced spatial resolution using normal compared to fine pitch, we recommend the use of normal pitch with 2 mm slice thickness to maximize image quality for ART while limiting scanning time. Only for sites for which improved CNR is required and reduced spatial resolution is acceptable, soft-tissue IR is recommended.

References Powered by Scopus

Fiji: An open-source platform for biological-image analysis

43564Citations
N/AReaders
Get full text

Accuracy of deformable image registration for contour propagation in adaptive lung radiotherapy

64Citations
N/AReaders
Get full text

On the dosimetric effect and reduction of inverse consistency and transitivity errors in deformable image registration for dose accumulation

55Citations
N/AReaders
Get full text

Cited by Powered by Scopus

AAPM Task Group Report 306: Quality control and assurance for tomotherapy: An update to Task Group Report 148

12Citations
N/AReaders
Get full text

Clinical Assessment of a Novel Ring Gantry Linear Accelerator-Mounted Helical Fan-Beam kVCT System

9Citations
N/AReaders
Get full text

Technical note: Characterization of novel iterative reconstructed cone beam CT images for dose tracking and adaptive radiotherapy on L-shape linacs

8Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Velten, C., Boyd, R., Jeong, K., Garg, M. K., & Tomé, W. A. (2020). Recommendations of megavoltage computed tomography settings for the implementation of adaptive radiotherapy on helical tomotherapy units. Journal of Applied Clinical Medical Physics, 21(5), 87–92. https://doi.org/10.1002/acm2.12859

Readers' Seniority

Tooltip

Researcher 2

67%

PhD / Post grad / Masters / Doc 1

33%

Readers' Discipline

Tooltip

Physics and Astronomy 2

67%

Medicine and Dentistry 1

33%

Save time finding and organizing research with Mendeley

Sign up for free