Measurement and Evaluation of Quantitative Performance of PET/CT Images before a Multicenter Clinical Trial /692/4017 /692/4028/67/2321 /123 article

11Citations
Citations of this article
22Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

To ensure the reliability of the planned multi-center clinical trial, we assessed the consistence and comparability of the quantitative parameters of the eight PET/CT units that will be used in this trial. PET/CT images were scanned using a PET NEMA image quality phantom (Biodex) on the eight units of Discovery PET/CT 690 from GE Healthcare. The scanning parameters were the same with the ones to be used in the planned trial. The 18F-NaF concentration in the background was 5.3 kBq/ml, while the ones in the spheres of diameter 37 mm, 22 mm, 17 mm and 10 mm were 8:1 as to that of the background and the ones in the spheres of diameter 28 mm and 13 mm were 0 kBq/ml. The consistency of hot sphere recovery coefficient (HRC), cold sphere recovery coefficient (CRC), hot sphere contrast (QH) and cold sphere contrast (Qc) among these 8 PET/CTs was analyzed. The variation of the main quantitative parameters of the eight PET/CT systems was within 10%, which is acceptable for the clinical trial.

Cite

CITATION STYLE

APA

Zhu, Y., Geng, C., Huang, J., Liu, J., Wu, N., Xin, J., … Geng, J. (2018). Measurement and Evaluation of Quantitative Performance of PET/CT Images before a Multicenter Clinical Trial /692/4017 /692/4028/67/2321 /123 article. Scientific Reports, 8(1). https://doi.org/10.1038/s41598-018-27143-4

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free