Background: The CloneSelect™ Imager system is an image-based visualisation system for cell growth assessment. Traditionally cell proliferation is measured with the colorimetric MTT assay.Results: Here we show that both the CloneSelect Imager and the MTT approach result in comparable EC50values when assaying the cytotoxicity of cisplatin and oxaliplatin on various cell lines. However, the image-based technique was found non-invasive, considerably quicker and more accurate than the MTT assay.Conclusions: This new image-based technique has the potential to replace the cumbersome MTT assay when fast, unbiased and high-throughput cytotoxicity assays are requested. © 2010 Marqués-Gallego et al; licensee BioMed Central Ltd.
CITATION STYLE
Marqués-Gallego, P., den Dulk, H., Backendorf, C., Brouwer, J., Reedijk, J., & Burke, J. F. (2010). Accurate non-invasive image-based cytotoxicity assays for cultured cells. BMC Biotechnology, 10. https://doi.org/10.1186/1472-6750-10-43
Mendeley helps you to discover research relevant for your work.