Abstract

A simple high-performance liquid chromatography method has been developed for determination of flurbiprofen in human plasma. The method was validated on an Ace C18 column using UV detection. The mobile phase was acetonitrile-0.05 M potassium dihydrogen phosphate solution (60:40, v/v) adjusted to pH 3.5 with phosphoric acid. The calibration curve was linear between the concentration range of 0.10-5.0 μg/mL. Intra- and inter-day precision values for flurbiprofen in plasma were <4.47, and accuracy (relative error) was better than 3.67%. The extraction recoveries of flurbiprofen from human plasma were between 93.0 and 98.9%. The limits of detection and quantification of flurbiprofen were 0.03 and 0.10 μg/mL, respectively. In addition, this assay was applied to determine the pharmacokinetic parameters of flurbiprofen in six healthy Turkish volunteers who had been given 100 mg flurbiprofen.

Cite

CITATION STYLE

APA

Yilmaz, B., & Erdem, A. F. (2015). Determination of flurbiprofen in human plasma by high-performance liquid chromatography. Journal of Chromatographic Science, 53(9), 1443–1448. https://doi.org/10.1093/chromsci/bmv032