A new prototype of an electronic jet-ventilator and its humidification system

12Citations
Citations of this article
19Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background: Adequate humidification in long-term jet ventilation is a critical aspect in terms of clinical safety. Aim: To assess a prototype of an electronic jet-ventilator and its humidification system. Methods: Forty patients with respiratory insufficiency were randomly allocated to one of four groups. The criterion for inclusion in this study was respiratory insufficiency exhibiting a Murray score above 2. The four groups of patients were ventilated with three different respirators and four different humidification systems. Patients in groups A and B received superimposed high-frequency jet ventilation (SHFJV) by an electronic jet-ventilator either with (group A) or without (group B) an additional humidification system. Patients in group C received high-frequency percussive ventilation (HFPV) by a pneumatic high-frequency respirator, using a hot water humidifier for warming and moistening the inspiration gas. Patients in group D received conventional mechanical ventilation using a standard intensive care unit respirator with a standard humidification system. SHFJV and HFPV were used for a period of 100 h (4 days). Results: A significantly low inspiration gas temperature was noted in patients in group B, initially (27.2 ± 2.5°C) and after 2 days (28.0 ± 1.6°C) (P<0.05). The percentage of relative humidity of the inspiration gas in patients in group B was also initially significantly low (69.8 ± 4.1%; P < 0.05) but rose to an average of 98 ± 2.8% after 2 h. The average percentage across all four groups amounted to 98 ± 0.4% after 2 h. Inflammation of the tracheal mucosa was found in patients in group B and the mucosal injury score (MIS) was significantly higher than in all the other groups. Patients in groups A, C and D showed no severe evidence of airway damage, exhibiting adequate values of relative humidity and temperature of the inspired gas. Conclusion: The problems of humidification associated with jet ventilation can be fully prevented by using this new jet-ventilator. These data were sustained by nondeteriorating MIS values at the end of the 4-day study period in groups A, C and D.

Cite

CITATION STYLE

APA

Kraincuk, P., Kepka, A., Ihra, G., Schabernig, C., & Aloy, A. (1999). A new prototype of an electronic jet-ventilator and its humidification system. Critical Care, 3(4), 101–110. https://doi.org/10.1186/cc351

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free