Comparison of esCCO and transthoracic echocardiography for non-invasive measurement of cardiac output intensive care

47Citations
Citations of this article
64Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

BackgroundThe esCCO monitor (ECG- estimated Continuous Cardiac Output, Nihon Kohden®) is a new non-invasive tool for estimating cardiac output (CO). It derives CO from the pulse wave transit time (PWTT) estimated by the ECG and the plethysmographic wave. An initial calibration is needed to refine the relation linking pulse pressure (measured by arterial pressure cuff) to PWTT. To assess the accuracy and reliability of the esCCO system, we performed an analysis of agreement of CO values obtained by transthoracic echocardiography (TTE).MethodsThirty-eight intensive care unit patients were prospectively included. CO was determined simultaneously using esCCO (CO esCCO) and TTE (COTTE) as our reference method.ResultsA total of 103 paired readings from 38 patients were collected. The correlation coefficient between COesCCO and COTTE was 0.61 (P<0.001). The Bland and Altman analysis corrected for repeated measures showed a bias of -1.6 litre min-1 and limits of agreement from -4.7 to 1.5 litre min-1, with a percentage error (2 sd/mean CO) of 49. The correlation for CO changes was significant (R0.63, P<0.001), but the concordance rate was poor (73). Polar plot analysis showed an angular bias of -9° with radial limits of agreement from -54° to 36°. The bias appeared to correlate with systemic vascular resistance (R-0.45, P<0.001).ConclusionsIn critically ill patients, the performance of the esCCO monitor was not clinically acceptable, and this monitor cannot be recommended in this setting. Moreover, the esCCO failed to trend CO data reliably. © 2012 The Author [2012].

Cite

CITATION STYLE

APA

Bataille, B., Bertuit, M., Mora, M., Mazerolles, M., Cocquet, P., Masson, B., … Larché, J. (2012). Comparison of esCCO and transthoracic echocardiography for non-invasive measurement of cardiac output intensive care. British Journal of Anaesthesia, 109(6), 879–886. https://doi.org/10.1093/bja/aes298

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free