The accurate and precise HPLC analytical method validated for the determination of Amlodipine besylate in pharmaceutical dosage form. The chromatographic separation is carried out on shimadzu HPLC system (LC-2010 CHT) with UV Vissible detector and C18(150mm x3.9 mm) 5 μm Column. The Mobile phase consists of Acetonitrile: Methanol: PH 3.0 Buffer (15 V: 35 V: 50 V) , at the flow rate of 1.0 ml/min and elutes were monitoring at 237 nm. The observed retention time for Amlodipine besylate was 12.3 min. The % RSD for system precision was 0.41 % and Method precision was 0.58 %. The method was found to linear (R=0.99996) in the Concentration range of 35-105 μg/ml (50 to 150%). The accuracy was in between 99.50-99.91%. Keywords: HPLC, Correlation coefficient, System suitability, Bias, % RSD and ICH guidelines
CITATION STYLE
Kumar, S., & Ram, B. (2019). Analytical method development and validation of Amlodipine besylate in tablet dosage form. Journal of Drug Delivery and Therapeutics, 9(4-A), 463–466. https://doi.org/10.22270/jddt.v9i4-a.3459