Method development and validation of cabozantinib by LC-MS/MS

2Citations
Citations of this article
7Readers
Mendeley users who have this article in their library.
Get full text

Abstract

The objective of this method is to be simple, precise, and economical performed by LC-MS/MS instrument. The mass spectrometric determination was performed using electrospray ionization in the positive mode with multiple reaction monitoring (MRM) mode and precursor to product ion transition to product ion of m/z 502.2 > 323 for cabozantinib. The effective separation of cabozantinib was achieved X-Bridge (2.1 mm × 100 mm, 3.5 μ) column and the mobile phase composition is 0.2% formic acid: acetonitrile (40:60 v/v), pumped at 0.12 ml/min flow rate. The Rt of cabozantinib was found to be 1.34 minutes. The LOD and LOQ were found at 1.5 ng/ml and 5 ng/ml concentrations and linearity concentrations were in a range of 5 ng/ml to 75 ng/ml with a regression correlation coefficient of 0.999. The % RSD value of accuracy was observed at 1.2–2.0. The marketed formulation assay was found to be 99.82%. The developed method and validation parameters were accepted as per USFDA guidelines

Cite

CITATION STYLE

APA

Venkateshwarlu, P., & Patel, M. M. (2022). Method development and validation of cabozantinib by LC-MS/MS. Pharmacia, 69(2), 407–413. https://doi.org/10.3897/pharmacia.69.e82684