Cardiovascular diseases such as acute myocardial infarction and heart failure accounted for the death of 17.5 million people (31% of all global deaths) in 2015. Monitoring the level of circulating N-terminal proBNP (NT-proBNP) is crucial for the detection of people at risk of heart failure. In this article, we describe a novel ultra-sensitive NT-proBNP test (us-NT-proBNP) that allows the quantification of circulating NT-proBNP in 30 min at 25 _C in the linear detection range of 7.0-600 pg/mL. It is a first report on the application of a fluorescence bead labeled detection antibody, DNA-guided detection method, and glass fiber membrane platform for the quantification of NT-proBNP in clinical samples. Limit of blank, limit of detection, and limit of quantification were 2.0 pg/mL, 3.7 pg/mL, and 7 pg/mL, respectively. The coefficient of variation was found to be less than 10% in the entire detection range of 7-600 pg/mL. The test demonstrated specificity for NT-proBNP without interferences from bilirubin, intra-lipid, biotin, and hemoglobin. The serial dilution test for plasma samples containing various NT-proBNP levels showed the linear decrement in concentration with the regression coefficient of 0.980-0.998. These results indicate that us-NT-proBNP test does not suffer from the interference of the plasma components for the measurement of NT-proBNP in clinical samples.