In SPRINT (Systolic Blood Pressure Intervention Trial), use of the Omron 907XL blood pressure (BP) monitor set at 5 minutes of antecedent rest to record BP produced an automated office BP value 7/6 mm Hg lower than awake ambulatory BP at 27 months. The authors studied the impact on automated office BP of setting the Omron 907XL to 0 minutes instead of 5 minutes of rest in patients with readings in the lower normal BP range, similar to on-treatment BP in the SPRINT intensive therapy group. Patients (n = 100) in cardiac rehabilitation were randomized to three BP readings at 1-minute intervals using an Omron 907XL BP device set for 5 or 0 minutes of antecedent rest. Mean (±standard deviation) automated office BP (mm Hg) after 5 minutes of rest (120.2 ± 14.6/66.9 ± 8.6 mm Hg) was lower (P
CITATION STYLE
Colella, T. J. F., Tahsinul, A., Gatto, H., Oh, P., & Myers, M. G. (2018). Antecedent rest may not be necessary for automated office blood pressure at lower treatment targets. Journal of Clinical Hypertension, 20(8), 1160–1164. https://doi.org/10.1111/jch.13319