Antecedent rest may not be necessary for automated office blood pressure at lower treatment targets

17Citations
Citations of this article
12Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

In SPRINT (Systolic Blood Pressure Intervention Trial), use of the Omron 907XL blood pressure (BP) monitor set at 5 minutes of antecedent rest to record BP produced an automated office BP value 7/6 mm Hg lower than awake ambulatory BP at 27 months. The authors studied the impact on automated office BP of setting the Omron 907XL to 0 minutes instead of 5 minutes of rest in patients with readings in the lower normal BP range, similar to on-treatment BP in the SPRINT intensive therapy group. Patients (n = 100) in cardiac rehabilitation were randomized to three BP readings at 1-minute intervals using an Omron 907XL BP device set for 5 or 0 minutes of antecedent rest. Mean (±standard deviation) automated office BP (mm Hg) after 5 minutes of rest (120.2 ± 14.6/66.9 ± 8.6 mm Hg) was lower (P

Cite

CITATION STYLE

APA

Colella, T. J. F., Tahsinul, A., Gatto, H., Oh, P., & Myers, M. G. (2018). Antecedent rest may not be necessary for automated office blood pressure at lower treatment targets. Journal of Clinical Hypertension, 20(8), 1160–1164. https://doi.org/10.1111/jch.13319