Numerical study of the respiCon sampler performance in the calm air

1Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Results of a numerical study of the RespiCon sampler performance in the calm air are presented. The air flow is described by the Navier-Stokes equations of axisymmetric stationary viscous flow of incompressible fluid that are numerically integrated by the computational fluid dynamics (CFD) software FLUENT. The collection efficiencies of RespiCon impactor stages agree quite well with experimental data and curves of the European standards for the thoracic and respirable dust fractions. The aspiration efficiencies derived from the numerical model overestimate the experimental data in the range of particle sizes of 10 μm < dp < 40 μm; however, they correctly predict the value of maximal size of aspirated particles. A new design of the RespiCon sampler with a higher volume flow rate was developed. Copyright © 2014 American Association for Aerosol Research.

Cite

CITATION STYLE

APA

Zaripov, S., & Koch, W. (2014). Numerical study of the respiCon sampler performance in the calm air. Aerosol Science and Technology, 48(1), 74–80. https://doi.org/10.1080/02786826.2013.859653