An in-vitro optical sensor designed to estimate Glycated Hemoglobin Levels

25Citations
Citations of this article
50Readers
Mendeley users who have this article in their library.

Abstract

The purpose of this research was to design an optical sensor for evaluating glycated hemoglobin (HbA1c) percentages in hemoglobin. The A1c sensors available in the market use invasive methods, while our device offers the possibility of non-invasive monitoring of HbA1c levels in diabetic patients. A prototype is assembled using two light emitting diodes with peak emission wavelengths of 535 nm and 593 nm, a photodiode, and a microcontroller. The proposed sensor measures the transmitted intensity in the form of an output voltage. We devise an approach to estimate the percentage of HbA1c in hemoglobin for a given solution. This estimation is based on the relative change in absorbance due to change in path length and molar absorption coefficients of hemoglobin and HbA1c, at the two wavelengths. We calculate the molar absorption coefficient of HbA1c at 535 nm and 593 nm wavelengths using the sensor, which is performed by a multiple variable regression analysis algorithm fed through the microcontroller. Specifically, the sensor output voltage with respect to the sample concentration is fitted to an exponentially decaying equation model. We used a commercial chemical assay called Control FD Glycohemoglobin A1c with known percentage HbA1c levels to verify our device measurements.

References Powered by Scopus

Calculation of protein extinction coefficients from amino acid sequence data

5305Citations
N/AReaders
Get full text

How to measure and predict the molar absorption coefficient of a protein

3659Citations
N/AReaders
Get full text

Spectroscopic Determination of Tryptophan and Tyrosine in Proteins

3108Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Wearable sensors enable personalized predictions of clinical laboratory measurements

160Citations
N/AReaders
Get full text

Analytical techniques for the detection of glycated haemoglobin underlining the sensors

36Citations
N/AReaders
Get full text

Derivation and validation of gray-box models to estimate noninvasive in-vivo percentage glycated hemoglobin using digital volume pulse waveform

23Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Mandal, S., & Manasreh, M. O. (2018). An in-vitro optical sensor designed to estimate Glycated Hemoglobin Levels. Sensors (Switzerland), 18(4). https://doi.org/10.3390/s18041084

Readers over time

‘18‘19‘20‘21‘22‘23‘24036912

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 16

70%

Professor / Associate Prof. 4

17%

Researcher 3

13%

Readers' Discipline

Tooltip

Engineering 8

47%

Chemistry 4

24%

Medicine and Dentistry 3

18%

Computer Science 2

12%

Save time finding and organizing research with Mendeley

Sign up for free
0