Introduction: Vasopressor infusions are essential in treating and preventing intraoperative hypotension. Closed-loop vasopressor therapy outperforms clinicians when the target is set at a mean arterial pressure (MAP) baseline, but little is known on the performance metrics of closed-loop vasopressor infusions when systolic arterial pressure (SAP) is the controlled variable. Methods: Patients undergoing intermediate- to high-risk abdominal surgery were included in this prospective cohort feasibility study. All patients received norepinephrine infusion through a computer controlled closed-loop system that targeted SAP at 130 mmHg. The primary objective was to determine the percent of case time in hypotension or under target defined as SAP below 10% of the target (SAP < 117 mmHg). Secondary objectives were the percent of case time “above target” (SAP > 10% of the target or >143 mmHg) and “in target” (within 10% of the SAP target or SAP between 117 and 143 mmHg). Results: A total of 12 patients were included. The closed-loop system infused norepinephrine for a median of 94.6% (25–75th percentile: 90.0–98.0%) of case time. The percentage of case time in hypotension or under target was only 1.8% (0.9–3.6%). The percentages of case time “above target” and “in target” were 4.7% (3.2–7.5%) and 92.4% (90.1–96.3%), respectively. Conclusions: This closed-loop vasopressor system minimizes intraoperative hypotension and maintains SAP within 10% of the target range for >90% of the case time in patients undergoing intermediate- to high-risk abdominal surgery.

Cite

CITATION STYLE

APA

Rinehart, J., Desebbe, O., Berna, A., Lam, I., Coeckelenbergh, S., Cannesson, M., & Joosten, A. (2022). Systolic Arterial Pressure Control Using an Automated Closed-Loop System for Vasopressor Infusion during Intermediate-to-High-Risk Surgery: A Feasibility Study. Journal of Personalized Medicine, 12(10). https://doi.org/10.3390/jpm12101554

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free