Position of effective electron source for shielded electron beams from a therapeutic linear accelerator

3Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.

Abstract

The effective electron source positions for the standard electron cones and for the shielded field sizes with cerrobend inserts were measured based on Inverse Square Law (ISL) and the Inverse Slope (IS) method for various electron energies. The charge measurements were carried out using a 0.6 cc ion chamber (PTW, Type 30001) connected to a PTW Unidos E digital electrometer in a polystyrene phantom for electron beam energies of 6-18 MeV. The resultant charge for 100 MU setting was measured at nominal source to surface distances (SSDs) of 100-120 cm for cone and cerrobend defined field sizes. The effective SSD (SSD eff) was found to be different for the same field size defined by electron applicator and the cerrobend shield placed in 25×25 cm standard cone. Strong dependency of SSD eff with field size and electron beam energy was noticed. The results from the ISL and IS method are consistent, hence either of the two methods can be used to determine the effective source position. Whenever treatment is to be given with shielded electron portal, the SSD eff for that field needs to be determined. Same SSD eff as that of the standard cone can be used for minimum shielded electron portals.

References Powered by Scopus

Clinical electron-beam dosimetry: Report of AAPM Radiation Therapy Committee Task Group No. 25

396Citations
N/AReaders
Get full text

Prediction of electron beam output factors

53Citations
N/AReaders
Get full text

Demonstration of relatively new electron dosimetry measurement techniques on the Mevatron 80

51Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Monte Carlo study on effective source to surface distance for electron beams from a mobile dedicated IORT accelerator

14Citations
N/AReaders
Get full text

Accuracy evaluation of distance inverse square law in determining virtual electron source location in Siemens Primus linac

8Citations
N/AReaders
Get full text

Determination of effective source to surface distance and cutout factor in small fields in electron beam radiotherapy: A comparison of different dosimeters

1Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Al Asmary, M. A., & Ravikumar, M. (2010). Position of effective electron source for shielded electron beams from a therapeutic linear accelerator. Polish Journal of Medical Physics and Engineering, 16(1), 11–21. https://doi.org/10.2478/v10013-010-0002-z

Readers over time

‘17‘18‘19‘20‘2100.250.50.751

Readers' Seniority

Tooltip

Researcher 2

67%

Professor / Associate Prof. 1

33%

Readers' Discipline

Tooltip

Medicine and Dentistry 2

50%

Physics and Astronomy 1

25%

Social Sciences 1

25%

Save time finding and organizing research with Mendeley

Sign up for free
0