Globally deployed COVID-19 fever screening devices using infrared thermographs consistently normalize high readings to afebrile range

  • Healy C
  • Segal Z
  • Hinnerichs C
  • et al.
5Citations
Citations of this article
29Readers
Mendeley users who have this article in their library.

Abstract

SIGNIFICANCE The need for regulatory review of infrared thermographs (IRTs) used on humans was removed in response to the unique circumstances of the SARS-CoV-2 pandemic (a.k.a., COVID-19). The market for these devices has since expanded considerably. This evaluation of IRT performance may have significant implications for febrility screening worldwide. AIM Perform controlled nonhuman trials of IRT devices to identify and quantify deviations in the human temperature range. APPROACH We compared IRT readings of a temperature-controlled non-human subject with one FDA-cleared IRT and one FDA-cleared handheld NCIT. In individual trials for each device, the subject was measured between 35°C and 40°C at 0.25°C increments. RESULTS The IRT device measurements were consistently normalized around the human mean (∼37  °  C). Temperatures were decremented as they approached the febrile range, and systematically reported as normal across all seven devices. This effect does not appear to be explained by a fixed offset or any known approach to estimating body temperature, or by random error. CONCLUSION The IRTs in this study generated human temperature measurements that were systematically biased to the mean human temperature. Given that these devices are utilized for sentinel detection of possible infectious disease transmission, and are now globally employed, the implications for reduced detection of febrility are a widespread false sense of security. This vulnerability must be considered with respect to facility access control, clinical applications, and travel screening in the context of the ongoing COVID-19 pandemic response.

Cite

CITATION STYLE

APA

Healy, C., Segal, Z., Hinnerichs, C., Ace, E., Ward, D., & Honovich, J. (2021). Globally deployed COVID-19 fever screening devices using infrared thermographs consistently normalize high readings to afebrile range. Journal of Biomedical Optics, 26(04). https://doi.org/10.1117/1.jbo.26.4.043009

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley