This study was aimed to establish a standard protocol and to quantitatively assess the reliability of harmonic analysis of the radial pulse wave measured by a harmonic wave analyzer (TD01C system). Both intraobserver and interobserver assessments were conducted to investigate whether the values of harmonics are stable in successive measurements. An intraclass correlation coefficient (ICC) and a Bland-Altman plot were used for this purpose. For the reliability assessments of the intraobserver and the interobserver, 22 subjects (mean age 45 ± 14 years; 14 males and 8 females) were enrolled. The first eleven harmonics of the radial pulse wave presented excellent repeatability ( ICCs >0.9 and p<0.001) for the intraobserver assessment and high reproducibility (ICCs range from 0.83 to 0.96 and p<0.001) for the interobserver assessment. The Bland-Altman plot indicated that more than 90% of harmonic values fell within two standard deviations of the mean difference. Thus, we concluded that the harmonic analysis of the radial pulse wave using the TD01C system is a feasible and reliable method to assess a hemodynamic characteristic in clinical trial.
CITATION STYLE
Chang, C. W., Chen, J. M., & Wang, W. K. (2015). Development of a Standard Protocol for the Harmonic Analysis of Radial Pulse Wave and Assessing Its Reliability in Healthy Humans. IEEE Journal of Translational Engineering in Health and Medicine, 3. https://doi.org/10.1109/JTEHM.2015.2490661