Comparison or recirculation percentage of the palindrome catheter and standard hemodialysis catheters in a swine model

42Citations
Citations of this article
21Readers
Mendeley users who have this article in their library.
Get full text

Abstract

PURPOSE: The purpose of this siudy was to evaluate in vivo recirculation of a new bi-directional dialysis catheter with a symmetric tip design and compare it to currently marketed staggered and split tip catheters. MATERIALS AND METHODS: Twelve catheters were placed in three 70-kg male swine. The catheters were connected to a roller type perista tic pump. Blood flow and recirculation rates were measured with ultrasound (US) dilution. Recirculation was measured in blood flow rates of 300 mL/min, 400 mL/min, and 500 mL/min. The dialysis lines were then reversed and recirculation was measured in flow rate of 400 mL/min. Measurements were conducted with the catheter tip in the right atrium (RA) and repeated with the catheter tip in the superior vena cava (SVC). RESULTS: Recirculation rates measured with the Palindrome catheter tip (Tyco Healthcare Group LP, Mansfield, MA) in the SVC and in the RA were 0%. When the dialysis lines were reversed, the Palindrome catheter average recirculation was 0% with the tip in the SVC and 1% (0%-3%) with the tip in the RA. The average recirculation rates measured with the split tip cathetur were 1% (0%-10%) in the SVC and 0% when the catheter tip was in the RA. When the dialysis lines were reversed split tip catheter recirculation was 11% (9%-16%) with the tip in the SVC and 13% (7%-18%) with the tip in the RA. With the staggered tip catheter, 1% (0%-5%) recirculation was measured. When the dialysis lines were reversed, the average recirculation in the staggered tip catheter measured 23% (18%-30%). Recirculation rate with reversal of dialysis lines was significantly lower with the Palindrome catheter when compared to the split tip and staggered tip designs (P < .05). CONCLUSIONS: Preliminary results in vivo indicate that the Palindrome catheter demonstrates minimal recirculation with reversal of the dialysis lines compared to currently marketed split tip and staggered tip catheters. © SIR, 2005.

Cite

CITATION STYLE

APA

Tal, M. G. (2005). Comparison or recirculation percentage of the palindrome catheter and standard hemodialysis catheters in a swine model. Journal of Vascular and Interventional Radiology, 16(9), 1237–1240. https://doi.org/10.1097/01.RVI.0000171700.45582.9E

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free