Purpose: To compare cumulative dissipated energy between two phacoemulsification machines. Setting: An ambulatory surgical center, Honolulu, Hawaii, USA. Design: Retrospective chart review. Methods: A total of 2,077 consecutive cases of cataract extraction by phacoemulsification performed by five surgeons from November 2012 to November 2014 were included in the study; 1,021 consecutive cases were performed using the Infiniti Vision System, followed by 1,056 consecutive cases performed using the Centurion Vision System. Results: The Centurion phacoemulsification system required less energy to remove a cataractous lens with an adjusted average energy reduction of 38% (5.09 percent-seconds) (P<0.001) across all surgeons in comparison to the Infiniti phacoemulsification system. The reduction in cumulative dissipated energy was statistically significant for each surgeon, with a range of 29%–45% (2.25–12.54 percent-seconds) (P=0.005-<0.001). Cumulative dissipated energy for both the Infiniti and Centurion systems varied directly with patient age, increasing an average of 2.38 percent-seconds/10 years. Conclusion: The Centurion phacoemulsification system required less energy to remove a cataractous lens in comparison to the Infiniti phacoemulsification system.
CITATION STYLE
Chen, M., Anderson, E., Hill, G., Chen, J. J., & Patrianakos, T. (2015). Comparison of cumulative dissipated energy between the Infiniti and Centurion phacoemulsification systems. Clinical Ophthalmology, 9, 1367–1372. https://doi.org/10.2147/OPTH.S88225