Comparison of cumulative dissipated energy between the Infiniti and Centurion phacoemulsification systems

46Citations
Citations of this article
36Readers
Mendeley users who have this article in their library.

Abstract

Purpose: To compare cumulative dissipated energy between two phacoemulsification machines. Setting: An ambulatory surgical center, Honolulu, Hawaii, USA. Design: Retrospective chart review. Methods: A total of 2,077 consecutive cases of cataract extraction by phacoemulsification performed by five surgeons from November 2012 to November 2014 were included in the study; 1,021 consecutive cases were performed using the Infiniti Vision System, followed by 1,056 consecutive cases performed using the Centurion Vision System. Results: The Centurion phacoemulsification system required less energy to remove a cataractous lens with an adjusted average energy reduction of 38% (5.09 percent-seconds) (P<0.001) across all surgeons in comparison to the Infiniti phacoemulsification system. The reduction in cumulative dissipated energy was statistically significant for each surgeon, with a range of 29%–45% (2.25–12.54 percent-seconds) (P=0.005-<0.001). Cumulative dissipated energy for both the Infiniti and Centurion systems varied directly with patient age, increasing an average of 2.38 percent-seconds/10 years. Conclusion: The Centurion phacoemulsification system required less energy to remove a cataractous lens in comparison to the Infiniti phacoemulsification system.

Cite

CITATION STYLE

APA

Chen, M., Anderson, E., Hill, G., Chen, J. J., & Patrianakos, T. (2015). Comparison of cumulative dissipated energy between the Infiniti and Centurion phacoemulsification systems. Clinical Ophthalmology, 9, 1367–1372. https://doi.org/10.2147/OPTH.S88225

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free