Design, modeling, and evaluation of the Eddy current sensor deeply implanted in the human body

5Citations
Citations of this article
18Readers
Mendeley users who have this article in their library.

Abstract

Joint replacement surgeries have enabled motion for millions of people suffering from arthritis or grave injuries. However, over 10% of these surgeries are revision surgeries. We have first analyzed the data from the worldwide orthopedic registers and concluded that the micromotion of orthopedic implants is the major reason for revisions. Then, we propose the use of inductive eddy current sensors for in vivo micromotion detection of the order of tens of µm. To design and evaluate its characteristics, we have developed efficient strategies for the accurate numerical simulation of eddy current sensors implanted in the human body. We present the response of the eddy current sensor as a function of its frequency and position based on the robust curve fit analysis. Sensitivity and Sensitivity Range parameters are defined for the present context and are evaluated. The proposed sensors are fabricated and tested in the bovine leg.

Cite

CITATION STYLE

APA

Khokle, R. P., Esselle, K. P., & Bokor, D. J. (2018). Design, modeling, and evaluation of the Eddy current sensor deeply implanted in the human body. Sensors (Switzerland), 18(11). https://doi.org/10.3390/s18113888

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley