We have assessed the potential clinical benefit of a new echo-Doppler device (Dynemo 3000) which provides a continuous measure of aortic blood flow (ABF) using an aortic flowmeter and a paediatric oesophageal probe, during repair of craniosynostosis in infants under general anaesthesia. The data recorded included: ABFi (i = indexed to body surface area), stroke volume (SVi), systemic vascular resistance (TSVRi), preejection period (PEP), left ventricular ejection time (LVET), mean arterial pressure (MAP), heart rate (HR) and central venous pressure (CVP). Data were collected: before (T1) and 3 min after skin incision (T2), at the time of maximal haemorrhage (T3) and at the end of the procedure (T4). Twelve infants (aged 7.0 (range 6-12) months) were included. ABFi, MAP and CVP were significantly lower at T3 compared with T1 (2.0 (0.8) vs 3.0 (0.8) litre min-1 m-2, 46.1 (5.8) vs 65.2 (8.9) mmHg and 2.8 (1.6) vs 5.2 (2.1) mmHg; P < 0.05). PEP/LVET ratio was significantly lower at T2 compared with T1 (0.25 (0.05) vs 0.30 (0.06)) and increased at T4 (0.36 (0.04); P < 0.05). These preliminary results suggest that this noninvasive ABF echo-Doppler device may be useful for continuous haemodynamic monitoring during a surgical procedure associated with haemorrhage in infants.
CITATION STYLE
Orliaguet, G. A., Meyer, P. G., Blanot, S., Jarreau, M. M., Charron, B., Cuttaree, H., … Renier, D. (1998). Non-invasive aortic blood flow measurement in infants during repair of craniosynostosis. British Journal of Anaesthesia, 81(5), 696–701. https://doi.org/10.1093/bja/81.5.696