Monte Carlo simulation of intrinsic count rate performance for scintillation gamma camera

0Citations
Citations of this article
4Readers
Mendeley users who have this article in their library.
Get full text

Abstract

This paper describes Monte Carlo simulation of intrinsic count rate performance of a scintillation gamma camera. MCNP Monte Carlo code was employed to calculate pulse height spectrum and detector efficiency. A custom code written in FORTRAN language was then developed to simulate, by Monte Carlo method, the distortion in pulse height spectrum due to the pile-up effect for paralyzable and non-paralyzable systems. The results of the simulations, compared with the experimental measurement of count rate performance, showed a good agreement between the two different approaches.

Cite

CITATION STYLE

APA

Mowlavi, A. A., De Denaro, M., Fornasier, M. R., & Binesh, A. (2009). Monte Carlo simulation of intrinsic count rate performance for scintillation gamma camera. In IFMBE Proceedings (Vol. 25, pp. 209–211). Springer Verlag. https://doi.org/10.1007/978-3-642-03882-2_55