Performance variability of seven commonly used self-monitoring of blood glucose systems: Clinical considerations for patients and providers

65Citations
Citations of this article
25Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background: Blood glucose data are frequently used in clinical decision making, thus it is critical that self-monitoring of blood glucose (SMBG) systems consistently provide accurate results. Concerns about SMBG accuracy have prompted the development of newly proposed International Organization for Standardization (ISO) standards: ≥95% of individual glucose results shall fall within ±15 mg/dl of the results of the manufacturer's reference procedure at glucose concentrations <100 mg/dl and within ±15% for values ≥100 mg/dl. We evaluated seven marketed systems against the current and proposed ISO criteria (criterion A). Method: Capillary blood samples were collected from 100 subjects and tested on seven systems: Accu-Chek Aviva Plus, Advocate Redi-Code, Element, Embrace, Prodigy Voice, TRUEbalance, and WaveSense Presto. Results were compared with manufacturer's documented reference system, YSI or perchloric acid hexokinase; three different strip lots from each system were tested on each subject, in duplicate. Results: Compared against current ISO criteria (≥95% within ±15 mg/dl for values <75 mg/dl and ±20% for values ≥75 mg/dl) the Accu-Chek Aviva Plus, Element, and WaveSense Presto systems met accuracy criteria. However, only the Accu-Chek Aviva Plus met the proposed ISO criteria (criterion A) in all three lots. The other six systems failed to meet the criteria in at least two of the three lots, showing lot-to-lot variability, high/low bias, and variations due to hematocrit. Conclusions: Inaccurate SMBG readings can potentially adversely impact clinical decision making and outcomes. Clinicians can reduce controllable variables by prescribing accurate SMBG systems. Adherence to the proposed ISO criteria should enhance patient safety by improving the accuracy of SMBG systems. © Diabetes Technology Society.

References Powered by Scopus

The Effect of Intensive Treatment of Diabetes on the Development and Progression of Long-Term Complications in Insulin-Dependent Diabetes Mellitus

0
23920Citations
N/AReaders
Get full text

The effects of normal as compared with low hematocrit values in patients with cardiac disease who are receiving hemodialysis and epoetin

1993Citations
N/AReaders
Get full text

World Medical Association Declaration of Helsinki: Recommendations Guiding Physicians in Biomedical Research Involving Human Subjects

0
1438Citations
N/AReaders
Get full text

Cited by Powered by Scopus

7. Diabetes Technology: Standards of Care in Diabetes—2023

314Citations
N/AReaders
Get full text

7. Diabetes Technology: Standards of Medical Care in Diabetes—2022

0
205Citations
N/AReaders
Get full text

Diabetes technology: Standards of medical care in diabetes−2021

0
172Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Brazg, R. L., Klaff, L. J., & Parkin, C. G. (2013). Performance variability of seven commonly used self-monitoring of blood glucose systems: Clinical considerations for patients and providers. Journal of Diabetes Science and Technology, 7(1), 144–152. https://doi.org/10.1177/193229681300700117

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 5

45%

Professor / Associate Prof. 3

27%

Researcher 3

27%

Readers' Discipline

Tooltip

Medicine and Dentistry 7

54%

Nursing and Health Professions 2

15%

Engineering 2

15%

Agricultural and Biological Sciences 2

15%

Save time finding and organizing research with Mendeley

Sign up for free