The research paper proposes a novel denoising method to improve the outcome of heart-sound (HS)-based heart-condition identification by applying the dual-tree complex wavelet transform (DTCWT) together with the adaptive neuro-fuzzy inference System (ANFIS) classifier. The method consists of three steps: first, preprocessing to eliminate 50 Hz noise; second, applying four successive levels of DTCWT to denoise and reconstruct the time-domain HS signal; third, to evaluate ANFIS on a total of 2735 HS recordings from an international dataset (PhysioNet Challenge 2016). The results show that the signal-to-noise ratio (SNR) with DTCWT was significantly improved (p < 0.001) as compared to original HS recordings. Quantitatively, there was an 11% to many decibel (dB)-fold increase in SNR after DTCWT, representing a significant improvement in denoising HS. In addition, the ANFIS, using six time-domain features, resulted in 55–86% precision, 51–98% recall, 53–86% f-score, and 54–86% MAcc compared to other attempts on the same dataset. Therefore, DTCWT is a successful technique in removing noise from biosignals such as HS recordings. The adaptive property of ANFIS exhibited capability in classifying HS recordings.
Mendeley helps you to discover research relevant for your work.
CITATION STYLE
Al-Naami, B., Fraihat, H., Al-Nabulsi, J., Gharaibeh, N. Y., Visconti, P., & Al-Hinnawi, A. R. (2022). Assessment of Dual-Tree Complex Wavelet Transform to Improve SNR in Collaboration with Neuro-Fuzzy System for Heart-Sound Identification. Electronics (Switzerland), 11(6). https://doi.org/10.3390/electronics11060938