Interface stresses in the resurfaced hip: Finite element analysis of load transmission in the femoral head

54Citations
Citations of this article
24Readers
Mendeley users who have this article in their library.

Abstract

The load transmission and interface stresses in the Wagner resurfaced femoral head were evaluated for the purpose of studying possible failure mechanisms. We found that unnatural stress patterns occur in the head and at the implant-bone interfaces, in addition to regions of stress protection in the bone, possibly enhancing interface failure and bone remodelling. However, these stresses are not higher than those reported for other kinds of prostheses, e.g. acetabular cup, tibial plateau. from these findings, together with clinical observations, it is hypothesized that the femoral surface cup is more sensitive to local loosening than other prostheses. This hypothesis would indicate that prosthetic designs should be analysed relative to their potential to provoke failure propagation, rather than only initiation of mechanical failure and loosening. © 1985 Informa UK Ltd All rights reserved: reproduction in whole or part not permitted.

Cite

CITATION STYLE

APA

Huiskes, R., Strens, P. H. G. E., van Heck, J., & Slooff, T. J. J. H. (1985). Interface stresses in the resurfaced hip: Finite element analysis of load transmission in the femoral head. Acta Orthopaedica, 56(6), 474–478. https://doi.org/10.3109/17453678508993038