Validated RP-HPLC Method for the Determination of Buspirone in Pharmaceutical Formulations

  • Basaveswara Rao M
  • Nagendrakumar A
  • Maiti S
  • et al.
N/ACitations
Citations of this article
5Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

A simple, selective, linear, precise, and accurate RP-HPLC method was developed and validated for the rapid assay of the Buspirone in tablet dosage form. Isocratic elution at a flow rate of 1.0 mL/min was employed on a symmetry C18 ( 250 × 4.6 mm, 5 μm in particle size) at ambient temperature. The mobile phase consisted of water : acetonitrile : methanol 45 : 35 : 20(V/V). The UV detection wavelength was 210 nm, and 20 μL sample was injected. The retention time for the Buspirone was 7.057 min. The percentage RSD for precision and accuracy of the method was found to be less than 2%. The method was validated as per the ICH guidelines. The method can be successfully applied for the routine analysis of Buspirone in tablet dosage form.

Cite

CITATION STYLE

APA

Basaveswara Rao, M. V., Nagendrakumar, A. V. D., Maiti, S., & Raja, G. (2011). Validated RP-HPLC Method for the Determination of Buspirone in Pharmaceutical Formulations. Chromatography Research International, 2011, 1–3. https://doi.org/10.4061/2011/232505