Validation of the BPLab® 24-hour blood pressure monitoring system according to the European standard BS EN 1060-4:2004 and British Hypertension Society protocol

20Citations
Citations of this article
18Readers
Mendeley users who have this article in their library.

Abstract

Introduction: Automatic blood pressure (BP) measuring devices are more and more often used in BP self-checks and in 24-hour BP monitoring. Nowadays, 24-hour BP monitoring is a necessary procedure in arterial hypertension treatment. The aim of this study was to validate the BPLab® ambulatory blood pressure monitor according to the European standard BS EN 1060-4:2004 and the British Hypertension Society (BHS) protocol, as well as to work out solutions regarding the suitability of using this device in clinical practice. Methods: A group of 85 patients of both sexes and different ages, who voluntarily agreed to take part in the tests and were given detailed instructions on the measurement technique were recruited for this study. The results of the BP measurement obtained by a qualifed operator using the BPLab® device were compared with the BP values measured using the Korotkov aus-cultatory method. Data were obtained simultaneously by two experts with experience of over 10 years and had completed a noninvasive BP measurement standardization training course. Discrepancies in the systolic and diastolic BP measurements (N = 510; 255 for each expert) were analyzed according to the criteria specifed in the BHS-93 protocol. Results: The device passed the requirements of the European Standard BS EN 1060-4:2004 and was graded 'A' according to the criteria of the BHS protocol for both systolic BP and diastolic B P. Conclusion: The BPLab® 24-hour ambulatory blood pressure monitoring device may be recommended for extensive clinical use. © 2011 Koudryavtcev and Lazarev, publisher and licensee Dove Medical Press Ltd.

Cite

CITATION STYLE

APA

Koudryavtcev, S. A., & Lazarev, V. M. (2011, October 26). Validation of the BPLab® 24-hour blood pressure monitoring system according to the European standard BS EN 1060-4:2004 and British Hypertension Society protocol. Medical Devices: Evidence and Research. Dove Medical Press Ltd. https://doi.org/10.2147/MDER.S25800

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free