Effects of External Electrical and Magnetic Fields on Pacemakers and Defibrillators

  • Beinart R
  • Nazarian S
N/ACitations
Citations of this article
13Readers
Mendeley users who have this article in their library.

Abstract

The overall risk of clinically significant adverse events related to EMI in recipients of CIEDs is very low. Therefore, no special precautions are needed when household appliances are used. Environmental and industrial sources of EMI are relatively safe when the exposure time is limited and distance from the CIEDs is maximized. The risk of EMI-induced events is highest within the hospital environment. Physician awareness of the possible interactions and methods to minimize them is warranted.© 2013 American Heart Association, Inc.

Cite

CITATION STYLE

APA

Beinart, R., & Nazarian, S. (2013). Effects of External Electrical and Magnetic Fields on Pacemakers and Defibrillators. Circulation, 128(25), 2799–2809. https://doi.org/10.1161/circulationaha.113.005697