Determination of cardiac output by an angle and diameter independent dual beam Doppler technique in critically ill infants

9Citations
Citations of this article
11Readers
Mendeley users who have this article in their library.

Abstract

Objective - To compare cardiac output measurements in critically ill infants by the dual beam Doppler and thermodilution techniques. Design - Prospective direct comparison of the two techniques. For statistical evaluation one randomly assigned paired measurement of every patient was used. Setting - Paediatric intensive care unit in a university hospital. Patients - 18 infants after open heart surgery aged 4-25 months (weight 4-10 kg). Interventions - Cardiac output measurements by dual beam Doppler and thermodilution techniques were performed within 10 minutes of each other and without knowledge of the results of the other methods. Multiple measurements were performed on some patients with a pharmacological or electrophysiological intervention or with a minimum of six hours between each pair of measurements. Measurements and main results - Three patients were excluded because of an inadequate Doppler signal or a significant residual shunt. Cardiac output measurements ranged from 0.4 to 2.2 l/min for the thermodilution technique and from 0.5 to 2.1 l/min for the dual beam Doppler technique. Agreement between both methods was acceptable. The mean difference between the two methods was 0.026 l/min with two standard deviations ranging from -0.20 to 0.26 l/min. Conclusion - The dual beam Doppler technique was shown to have promise for the non-invasive determination of cardiac output in critically ill infants.

References Powered by Scopus

STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT

42119Citations
N/AReaders
Get full text

Noninvasive Doppler determination of cardiac output in man. Clinical validation

501Citations
N/AReaders
Get full text

Noninvasive pulsed Doppler determination of cardiac output in neonates and children

207Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Accuracy and repeatability of pediatric cardiac output measurement using Doppler: 20-Year review of the literature

131Citations
N/AReaders
Get full text

Electrical velocimetry for measuring cardiac output in children with congenital heart disease

113Citations
N/AReaders
Get full text

Accuracy and precision of minimally-invasive cardiac output monitoring in children: a systematic review and meta-analysis

59Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Wippermann, C. F., Schranz, D., Huth, R., Zepp, F., Oelert, H., & Jungst, B. K. (1992). Determination of cardiac output by an angle and diameter independent dual beam Doppler technique in critically ill infants. British Heart Journal, 67(2), 180–184. https://doi.org/10.1136/hrt.67.2.180

Readers over time

‘14‘16‘17‘19‘20‘21‘2501234

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 2

40%

Professor / Associate Prof. 1

20%

Lecturer / Post doc 1

20%

Researcher 1

20%

Readers' Discipline

Tooltip

Medicine and Dentistry 5

50%

Engineering 4

40%

Nursing and Health Professions 1

10%

Save time finding and organizing research with Mendeley

Sign up for free
0