The aim of this study was to investigate whether FTIR spectroscopy is an accurate and valid technique for the assessment of quality parameters in raw cow's milk: fat, protein, lactose, and total solids. The assessment was based on calibration series and comparison with reference material. Furthermore, it takes into account the results obtained in the inter-laboratory comparisons (proficiency testing). The calibration samples were purchased from the accredited regional reference laboratories. The validation parameters included linearity, accuracy, repeatability, reproducibility, and robustness. The linearity ratio was 0.95%. The biases calculated for the fat, protein, lactose and dry matter were -0.33, 0.31, -0.25, and 0.06 respectively. The F value from the F-test was used to determine the significant differences between two independent sets of the results. The obtained results were as follows: 1.469 for fat, 1.634 for protein, 1.192 for lactose, and 0.528 for dry matter. The intra-laboratory reproducibility calculated as the Horwitz Ratios for all parameters were within the criterion limits (0.5 to 0.8). The data obtained for carry-over were 0.27% for fat, 0.52% for protein, 0.47% for lactose, and 0.47% for dry matter. Based on the obtained results it can be concluded that the FTIR spectroscopy is a reliable instrumental technique for the determination of fat, protein, lactose and total solids in raw cow's milk.
CITATION STYLE
Kučević, D. S., Glamočić, D. M., Trivunović, S. J., Jajić, I. M., Perišić, B., & Krstović, S. Z. (2017). Validation and application of FTIR spectroscopy in raw milk analysis. Acta Periodica Technologica, 48, 167–175. https://doi.org/10.2298/APT1748167K