Validation of QCL CF4 Gas Analyzer for Sensitivity and Selectivity

1Citations
Citations of this article
1Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Quantum cascade lasers (QCL) has open the mid-infrared (mid-IR) spectral range for laser gas monitors. For perfluorocarbon gases (PFCs) such as CF4, continuous emission monitors are now commercially available. Since these instruments can only make use of a narrow IR band, spectral interference from other gas species is not easily resolved. In this paper a commercial QCL CF4 analyser is validated with respect to sensitivity and selectivity. In the laboratory, the laser was mounted on a 0.3-m measurement cell. A gas mixer provided humidified gas mixtures of CF4 and methane. The results indicated that while there was no interference from water in the range up to 8000 ppm H2O, a small interference from methane was observed. This was quantified to be approximately 3 ppb CF4 per ppm of methane. The sensitivity of the instrument was found to be 10.5 ppb CF4 for one-meter light path.

Cite

CITATION STYLE

APA

Aarhaug, T. A. (2020). Validation of QCL CF4 Gas Analyzer for Sensitivity and Selectivity. In Minerals, Metals and Materials Series (pp. 744–750). Springer Science and Business Media Deutschland GmbH. https://doi.org/10.1007/978-3-030-36408-3_100