Continuous ambulatory haemodynamic monitoring with an implantable system. The feasibility of a new technique

38Citations
Citations of this article
11Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Aims. This study evaluates the feasibility and safety of a completely implantable system for long-term ambulatory monitoring of important haemodynamic parameters in patients with severe cardiopulmonary disease. Methods. The design of the implantable monitoring system is similar to a conventional single lead pacemaker. A lead with incorporated biosensors for the continuous recording of pressure and oxygen saturation signals is positioned in the right ventricle and connected to a monitor and memory device subcutaneously implanted like an ordinary pacemaker can. Results. Five patients with implanted haemodynamic monitoring systems have been followed for from 7 to 16 months. Continuous measurements of activity, heart rate, mixed venous oxygen saturation and estimated pulmonary artery diastolic pressure were registered with variable resolution during daily living and predefined provocations. The memory covered a maximum of 3 weeks at low resolution. The monitored parameters showed an adequate and significant response to various haemodynamic situations. Except for the demand of recalibration of two oxygen sensors, there were no technical problems and the quality of data were excellent. Conclusion. Long-term ambulatory haemodynamic monitoring is feasible and potentially useful for the management of patients with severe cardiopulmonary disease.

Cited by Powered by Scopus

Ongoing right ventricular hemodynamics in heart failure: Clinical value of measurements derived from an implantable monitoring system

317Citations
N/AReaders
Get full text

Continuous ambulatory right heart pressure measurements with an implantable hemodynamic monitor: A multicenter, 12-month follow-up study of patients with chronic heart failure

157Citations
N/AReaders
Get full text

Continuous ambulatory monitoring of absolute right ventricular pressure and mixed venous oxygen saturation in patients with heart failure using an implantable haemodynamic monitor: Results of a 1 year multicentre feasibility study

76Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Ohlsson, A., Nordlander, R., Bennett, T., Bitkover, C., Kjellström, B., Lee, B., & Rydén, L. (1998). Continuous ambulatory haemodynamic monitoring with an implantable system. The feasibility of a new technique. European Heart Journal, 19(1), 174–184. https://doi.org/10.1053/euhj.1997.0563

Readers' Seniority

Tooltip

Researcher 4

50%

PhD / Post grad / Masters / Doc 3

38%

Professor / Associate Prof. 1

13%

Readers' Discipline

Tooltip

Medicine and Dentistry 4

50%

Nursing and Health Professions 2

25%

Sports and Recreations 1

13%

Engineering 1

13%

Save time finding and organizing research with Mendeley

Sign up for free