Abstract

Aims: To test recommended implantable cardioverter defibrillator (ICD) follow-up methods by 'in-person evaluations' (IPE) vs. 'remote Home Monitoring' (HM). Methods and results: ICD patients were randomized 2:1 to automatic HM or to Conventional monitoring, with follow-up checks scheduled at 3, 6, 9, 12, and 15 months post-implant. Conventional patients were evaluated with IPE only. Home Monitoring patients were assessed remotely only for 1 year between 3 and 15 month evaluations. Adherence to follow-up was measured. HM and Conventional patients were similar (age 63 years, 72% male, left ventricular ejection fraction 29%, primary prevention 73%, DDD 57%). Conventional management suffered greater patient attrition during the trial (20.1 vs. 14.2% in HM, P = 0.007). Three month follow-up occurred in 84% in both groups. There was 100% adherence (5 of 5 checks) in 47.3% Conventional vs. 59.7% HM (P < 0.001). Between 3 and 15 months, HM exhibited superior (2.2×) adherence to scheduled follow-up [incidence of failed follow up was 146 of 2421 (6.0%) in HM vs. 145 of 1098 (13.2%) in Conventional, P < 0.001] and punctuality. In HM (daily transmission success rate median 91%), transmission loss caused only 22 of 2275 (0.97%) failed HM evaluations between 3 and 15 months; others resulted from clinic oversight. Overall IPE failure rate in Conventional [193 of 1841 (10.5%) exceeded that in HM [97 of 1484 (6.5%), P < 0.001] by 62%, i.e. HM patients remained more loyal to IPE when this was mandated. Conclusion: Automatic remote monitoring better preserves patient retention and adherence to scheduled follow-up compared with IPE. Clinical trial registration NCT00336284. © The Author 2014.

Cite

CITATION STYLE

APA

Varma, N., Michalski, J., Stambler, B., & Pavri, B. B. (2014). Superiority of automatic remote monitoring compared with in-person evaluation for scheduled ICD follow-up in the TRUST trial - Testing execution of the recommendations. European Heart Journal, 35(20), 1345–1352. https://doi.org/10.1093/eurheartj/ehu066

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free