Origins of Baseline Drift and Distortion in Fourier Transform Spectra

12Citations
Citations of this article
42Readers
Mendeley users who have this article in their library.

Abstract

The spectrum scanned by a Fourier transform spectrometer (FTIR) often has a baseline drift. However, baseline distortion rarely occurs in a laboratory owing to the insignificant effects of environmental vibrations and electromagnetic factors. Even if it occurs, the distorted spectrum can be manually eliminated. However, in a complex environment, especially after the long-term operation of a spectrometer, the scanned spectrum may be distorted to different degrees. Herein, the origins of spectral baseline drifts and distortions are analyzed and simulated using MATLAB; furthermore, a baseline correction method based on the baseline-type model is proposed. The results of experiments performed on the methane spectrum confirm that the proposed method outperformed the improved modified multi-polynomial fitting and iterative averaging methods.

Cite

CITATION STYLE

APA

Zhang, F., Tang, X., & Li, L. (2022). Origins of Baseline Drift and Distortion in Fourier Transform Spectra. Molecules, 27(13). https://doi.org/10.3390/molecules27134287