Comparison of ultrasonographic measurement of optic nerve sheath diameter (ONSD) versus direct measurement of intracranial pressure (ICP) in traumatic brain injury patients

9Citations
Citations of this article
17Readers
Mendeley users who have this article in their library.

Abstract

Aim. To compare ultrasonographic measurement of optic nerve sheath diameter (ONSD) with direct measurement of intra- cranial pressure (ICP) and evaluate the correlation between them in traumatic brain injury patients. Patients and methods. Twenty traumatic brain injury patients were enrolled in the study. Patients had a median Glasgow Coma Scale (GCS) score of 5. All patients underwent noninvasive measurement of ICP by ultrasonographic measurement of ONSD and invasive ICP measurement using an intraventricular catheter. We used MedCalc to analyze data. Results. We compared the ONSD measurement in patients with ICP less than 20 cm H2O with patients with ICP greater than 20 cm H2O. The mean ONSD for 8 patients with ICP > 20 cmH2O was 7.6 0.8 mm, and for 18 patients with ICP < 20 cm H2O was 5.6 0.6 mm. Student t-test comparison between these two groups showed a statistically significant mean difference of 1.99 (95 % CI = 1.4-2.6). The ROC curve showed that the optimal cutoff value of ONSD for predicting ICP > 20 cm H2O was 6.1, with sensitivity of 100% and specificity of 83 %. (the area under the curve was 0.98, 95 % CI = 0.825- 0.990). In this study, the usually used cutoff of ONSD > 5 mm gives a sensitivity of 100 %, but a specificity of only 22%. Conclusion. Ultrasonographic measurement of ONSD correlates with direct measurement of ICP. This method has potential as a screening test for elevated ICP in traumatic brain injury.

Figures

  • Figure 1. ROC curve for ONSD for patients with ICP > 20 cm H2O (threshold of ONSD = 6,1) ICP, intracranial pressure; ONSD, optic nerve sheath diameter; ROC, receiver operator curve.

Author supplied keywords

References Powered by Scopus

Correlation of optic nerve sheath diameter with direct measurement of intracranial pressure

433Citations
N/AReaders
Get full text

Ultrasonography of the optic nerve sheath may be useful for detecting raised intracranial pressure after severe brain injury

363Citations
N/AReaders
Get full text

Emergency Department Sonographic Measurement of Optic Nerve Sheath Diameter to Detect Findings of Increased Intracranial Pressure in Adult Head Injury Patients

338Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Bedside optic nerve ultrasonography for diagnosing increased intracranial pressure

74Citations
N/AReaders
Get full text

Neuromuscular ultrasound of cranial nerves

40Citations
N/AReaders
Get full text

Optic Nerve Sheath Diameter Ultrasound for Raised Intracranial Pressure: A Literature Review and Meta-analysis of its Diagnostic Accuracy

35Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Širanovi, M., Turkoví, T. M., Gopčevi, A., Mijo, K. I., Kovač, N., Kovač, J., … Vušiś, M. (2011). Comparison of ultrasonographic measurement of optic nerve sheath diameter (ONSD) versus direct measurement of intracranial pressure (ICP) in traumatic brain injury patients. Signa Vitae, 6(1), 33–35. https://doi.org/10.22514/sv61.042011.5

Readers over time

‘12‘13‘14‘16‘17‘20‘21‘22‘2401234

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 7

64%

Professor / Associate Prof. 2

18%

Researcher 2

18%

Readers' Discipline

Tooltip

Medicine and Dentistry 10

71%

Engineering 2

14%

Agricultural and Biological Sciences 1

7%

Neuroscience 1

7%

Save time finding and organizing research with Mendeley

Sign up for free
0