Abstract

This work is focused on the design and implementation of a microsystem for continuous colorimetric determination of cobalt in aqueous solutions in real time. The main goal is the colorimetric determination of cobalt by adapting colorimetric methodologies used in discontinuous analysis systems to a continuous flow system and its subsequent miniaturization using the LTCC (Low Temperature Co- fired Ceramics) technology. The measure absorbance was recorded using an optical detection system based on a LED (Light Emitting Diode) with a wavelength of 540 nm.

Cite

CITATION STYLE

APA

Bustos López, O. N., Perezgasga, F. V., Moreno Casillas, H. A., Chamarro, J. A., & Fuentes, H. C. (2014). Analytical microsystem for the monitoring and analysis of cobalt in aqueous solutions using LTCC technology. Journal of the Mexican Chemical Society, 58(4), 411–415. https://doi.org/10.29356/jmcs.v58i4.49