We have compared occupational exposure to isoflurane and desflurane during cardiopulmonary bypass, with and without a scavenging system at the membrane oxygenator outlet. Trace concentrations of volatile anaesthetics were measured by a direct reading instrument in 40 elective heart surgery procedures. Measurements were obtained in the breathing zones of the anaesthetist and perfusionist. When a scavenging system was used, median desflurane values were less than 0.3 ppm and isoflurane values less than 0.2 ppm. Without a scavenging system values were, in general, three(isoflurane) to five- (desflurane) fold higher. We conclude that the use of a scavenging system at the membrane oxygenator outlet can reduce occupational exposure to volatile anaesthetics. We therefore recommend routine use of scavenging devices during cardiopulmonary bypass.
CITATION STYLE
Hoerauf, K., Harth, M., Wild, K., & Hobbhahn, J. (1997). Occupational exposure to desflurane and isoflurane during cardiopulmonary bypass: Is the gas outlet of the membrane oxygenator an operating theatre pollution hazard? British Journal of Anaesthesia, 78(4), 378–380. https://doi.org/10.1093/bja/78.4.378