After osteosynthesis of the proximal humerus by K-wires, loosening and secondary loss can occur. This study tested primary fixation of intramedullayr nails made from a shape memory alloy (SMA) Nitinol® (NiTi), compared to conventional steel nails by rotational tests. We found that NiTi-nails have less rotational stability.
CITATION STYLE
Wiebking, U., Stapel, N., Hurschler, C., & Gösling, T. (2009). Rotation resistance of new intramedullary nail made form shape memory alloy depends upon the bone diameter. In IFMBE Proceedings (Vol. 25, pp. 91–94). Springer Verlag. https://doi.org/10.1007/978-3-642-03906-5_25