The most discriminating method that is currently applied in routine forensic elemental analysis of glass is laser ablation in combination with quadrupole or sector field based inductively coupled plasma mass spectrometry (LA-ICPMS) following the standard method ASTM E2927-16E1. Due to the sequential measurement method, sample size constraints of approximately 400 μm × 200 μm × 100 μm with six individual measurements have been reported to provide optimum parameters for glass discrimination. The necessary sample size is given by the spot size, number of laser pulses per replicate and number of replicate measurements. In this study, a single-pulse laser ablation inductively coupled plasma time of flight mass spectrometry (Single-Pulse LA-ICP-TOFMS) method was developed and applied for the matching and mismatching of forensic float glass evidence. 110 laser pulses with an ablation spot diameter of 90 μm were applied to float glass fragments of a size below 400 μm in diameter. 18 elements were quantified from each individual laser pulse in order to compare the corresponding concentrations between various glass fragments. In contrast to previous work, a modified 5-sigma criterion was used to successfully match fragments from the same source and mismatch fragments from differing sources. The data set generated within this study demonstrates new capabilities when using quasi simultaneous signal detection and reduced the necessary sample volume from 400 μm × 200 μm × 100 μm to 100 μm × 100 μm × 33 μm, which corresponds to a reduction in sample material from 20 μg to 0.8 μg. The proof of concept is shown for the application of a single-pulse based method for glass fragment analysis in a forensic context, which would allow the measurement of smaller fragments than previously possible.

References Powered by Scopus

Limits for Qualitative Detection and Quantitative Determination: Application to Radiochemistry

3869Citations
N/AReaders
Get full text

Determination of reference values for NIST SRM 610-617 glasses following ISO guidelines

1592Citations
N/AReaders
Get full text

Laser ablation inductively coupled plasma mass spectrometric transient signal data acquisition and analyte concentration calculation

1447Citations
N/AReaders
Get full text

Cited by Powered by Scopus

The addition of biochar as a fertilizer supplement for the attenuation of potentially toxic elements in phosphogypsum-amended soil

28Citations
N/AReaders
Get full text

Parallel flow ablation cell for short signal duration in LA-ICP-TOFMS element imaging

25Citations
N/AReaders
Get full text

Atomic spectrometry update: Review of advances in the analysis of metals, chemicals and materials

21Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Becker, P., Neff, C., Hess, S., Weis, P., & Günther, D. (2020). Forensic float glass fragment analysis using single-pulse laser ablation inductively coupled plasma time of flight mass spectrometry. Journal of Analytical Atomic Spectrometry, 35(10), 2248–2254. https://doi.org/10.1039/d0ja00284d

Readers over time

‘20‘21‘22‘23‘2402468

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 6

55%

Researcher 4

36%

Professor / Associate Prof. 1

9%

Readers' Discipline

Tooltip

Chemistry 6

60%

Engineering 2

20%

Chemical Engineering 1

10%

Materials Science 1

10%

Save time finding and organizing research with Mendeley

Sign up for free
0