A simple, reproducible and efficient method for the determination of metformin hydrochloride (MET) was developed and validated. The analysis complied with Beer's law in the concentration range of 8-13 μg/mL at 233 nm for MET. In our study the validation of analytical method for determination of MET by UV in tablets formulation was performed in accordance the parameters including-system suitability, specificity, limit of quantification, limit of detection, linearity of response, accuracy, precision (reproducibility & repeatability), robustness (change of wave length ±2 nm).
CITATION STYLE
Mishra, K., Soni, H., Nayak, G., Patel, S. S., & Singhai, A. K. (2011). Method development and validation of metformin hydrochloride in tablet dosage form. E-Journal of Chemistry, 8(3), 1309–1313. https://doi.org/10.1155/2011/768014