Abstract

Purpose: The Leksell Gamma Plan Convolution algorithm (LGP-Convolution) has not been widely adopted. This mainly stems from the higher calculated beam-on times relative to the standard ray tracing-based LGP-TMR10 dose calculation algorithm. This study aims to evaluate the accuracy of the LGP-Convolution in scenarios where the treated lesions are in the vicinity of or encompassed by bone and/or air inhomogeneities. Methods: The solid water dosimetry phantom provided by the vendor was modified to include bone and air inhomogeneities. Two treatment planning scenarios were investigated involving a single shot and multiple shots, respectively. Treatment planning and dose prescription were performed using the LGP-Convolution algorithm. Triple channel film dosimetry was performed using GafChromic EBT3 films calibrated in terms of absorbed dose to water in a 60Co beam. Monte Carlo (MC) simulation dosimetry was also performed in the inhomogeneous experimental geometry using the EGSnrc MC platform and a previously validated sector-based phase-space source model. MC simulations were also employed to determine correction factors required for converting EBT3 measurements at points within the bone and air inhomogeneities from dose-to-water values to the corresponding dose to medium values. Results and Conclusions: EBT3 dose to medium correction factors ranged with field size (4, 8, or 16 mm) within 0.941–0.946 for bone and 0.745–0.749 for air inhomogeneities. An excellent agreement was found between the LGP-Convolution calculations with corresponding EBT3 and MC dose to medium results at all measurement points, except those located inside the air inhomogeneity. The latter is of no clinical importance and excluding them yielded gamma index passing rates of nearly 100% for 3% local dose difference and 1 mm distance-to-agreement criteria. The excellent agreement observed between LGP-Convolution calculations and film as well as MC results of dose to medium indicates that the latter is the quantity reported by the LGP-Convolution.

References Powered by Scopus

Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 version

878Citations
N/AReaders
Get full text

Correlation between CT numbers and tissue parameters needed for Monte Carlo simulations of clinical dose distributions

683Citations
N/AReaders
Get full text

A new formalism for reference dosimetry of small and nonstandard fields

479Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Advanced External Beam Stereotactic Radiotherapy for Skull Base Reirradiation

0Citations
N/AReaders
Get full text

Adaptation of dose-prescription for vestibular schwannoma radiosurgery taking body contouring method and heterogeneous material into account

0Citations
N/AReaders
Get full text

Dose assessment by TL dosimetry according to parameters used in treatment plans for obsessive compulsive disorder using the Gamma Knife@ Perfexion™ device

0Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Pantelis, E., Logothetis, A., Zoros, E., Pappas, E. P., Papagiannis, P., Paddick, I., … Karaiskos, P. (2023). Dosimetric accuracy of the Convolution algorithm for Leksell Gamma Plan radiosurgery treatment planning: Evaluation in the presence of clinically relevant inhomogeneities. Journal of Applied Clinical Medical Physics, 24(5). https://doi.org/10.1002/acm2.13903

Readers' Seniority

Tooltip

Professor / Associate Prof. 3

43%

PhD / Post grad / Masters / Doc 2

29%

Researcher 2

29%

Readers' Discipline

Tooltip

Physics and Astronomy 4

57%

Medicine and Dentistry 2

29%

Agricultural and Biological Sciences 1

14%

Article Metrics

Tooltip
Mentions
News Mentions: 1

Save time finding and organizing research with Mendeley

Sign up for free