Finite element study on the micromotion of cementless total hip arthroplasty

0Citations
Citations of this article
9Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Major problem associated with the cementless design is the lack of stability. Several design features were introduced to provide the much needed stability for bone integration. These include fins, fiber mesh, beads, indentations and collars. The use of collar has been a subject of controversy with mixed reports of clinical outcomes. In this study, the effectiveness of collar in terms of providing stability was analysed using finite element method. Three dimensional model of a human femur was constructed from CT dataset and hip arthroplasty was simulated. With an applied joint contact load simulating physiological gait, micromotion of the implant relative to the bone was calculated via an experimentally validated algorithm. Results showed that collar did not improve stability of the whole stem but only around the medial calcar region. © 2010 International Federation for Medical and Biological Engineering.

Cite

CITATION STYLE

APA

Baharuddin, M. Y., & Kadir, M. R. A. (2010). Finite element study on the micromotion of cementless total hip arthroplasty. In IFMBE Proceedings (Vol. 31 IFMBE, pp. 605–607). https://doi.org/10.1007/978-3-642-14515-5_154