Interscan registration using navigator echoes

15Citations
Citations of this article
33Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

A common problem in clinical MRI is anatomic misalignment of imaging slices across successive examinations. This unnecessarily complicates the radiologic assessment of anatomic change over time on serial MRI studies. To address this problem, spherical navigator echoes, which can detect rigid body motion in all six degrees of freedom, were used to guide spatial location and orientation adjustments to an exam prescription to match the reference frame of images acquired in an earlier exam. An initial linear navigator echo is also necessary to effect coarse Z translation adjustments prior to fine six degrees of freedom adjustment with a spherical navigator echo. Results of this technique are presented for head image volumes of five volunteers. Each volunteer was imaged on two scanners. In all cases, the reference frame adjustments provided by the navigator echoes substantially improved the alignment of the latter exam and performed well compared to retrospective image-based registration. © 2004 Wiley-Liss, Inc.

Cite

CITATION STYLE

APA

Welch, E. B., Manduca, A., Grimm, R. C., & Jack, C. R. (2004). Interscan registration using navigator echoes. Magnetic Resonance in Medicine, 52(6), 1448–1452. https://doi.org/10.1002/mrm.20275