Transient elastography is unreliable for detection of cirrhosis in patients with acute liver damage

447Citations
Citations of this article
124Readers
Mendeley users who have this article in their library.

Abstract

Transient elastography [FibroScan (FS)] is a rapid, noninvasive, and reproducible method for measuring liver stiffness, which correlates with the degree of liver fibrosis in patients with chronic hepatitis. Whether FS is useful in the detection of preexisting liver fibrosis/cirrhosis in patients presenting with acute liver damage is unclear. Patients with acute liver damage of different etiologies were analyzed. Liver stiffness was measured during the acute phase of the liver damage and followed up to the end of the acute phase. A total of 20 patients were included in the study. In 15 of the 20 patients, initial liver stiffness values measured by FS during the acute phase of the liver damage were suggestive of liver cirrhosis. However, none of these 15 patients showed any signs of liver cirrhosis in the physical examination, ultrasound examination, or liver histology [performed in 11 of 15 (73%) patients]. A significant difference was observed in the initial bilirubin levels (5.8 ± 6.5 mg/dL versus 15.7 ± 11.8 mg/dL; P = 0.042) and age (32.4 ± 17.5 years versus 49.7 ± 15.8 years; P = 0.042) between patients with liver stiffness below or above 12.5 kPa. Six patients with initially high liver stiffness were followed up to abatement of the acute hepatitic phase; in all of them, liver stiffness values decreased to values below the cutoff value for liver cirrhosis. Conclusion: Transient elastography frequently yields pathologically high values in patients with acute liver damage and is unsuitable for detecting cirrhosis/fibrosis in these patients. Copyright © 2007 by the American Association for the Study of Liver Diseases.

References Powered by Scopus

Transient elastography: A new noninvasive method for assessment of hepatic fibrosis

2398Citations
N/AReaders
Get full text

Prospective comparison of transient elastography, Fibrotest, APRI, and liver biopsy for the assessment of fibrosis in chronic hepatitis C

2213Citations
N/AReaders
Get full text

Sampling Variability of Liver Fibrosis in Chronic Hepatitis C

2122Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Diagnosis, management, and treatment of hepatitis C: An update

2732Citations
N/AReaders
Get full text

EASL-ALEH Clinical Practice Guidelines: Non-invasive tests for evaluation of liver disease severity and prognosis

1570Citations
N/AReaders
Get full text

Non-invasive evaluation of liver fibrosis using transient elastography

1167Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Sagir, A., Erhardt, A., Schmitt, M., & Häussinger, D. (2008). Transient elastography is unreliable for detection of cirrhosis in patients with acute liver damage. Hepatology, 47(2), 592–595. https://doi.org/10.1002/hep.22056

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 39

51%

Researcher 24

31%

Professor / Associate Prof. 14

18%

Readers' Discipline

Tooltip

Medicine and Dentistry 62

74%

Engineering 14

17%

Biochemistry, Genetics and Molecular Bi... 5

6%

Physics and Astronomy 3

4%

Save time finding and organizing research with Mendeley

Sign up for free