International federation of clinical chemistry standardization project for the measurement of lipoprotein(a). Phase I. Evaluation of the analytical performance of lipoprotein(a) assay systems and commercial calibrators

66Citations
Citations of this article
19Readers
Mendeley users who have this article in their library.
Get full text

Abstract

A secondary reference material for lipoprotein(a) is required to standardize the measurement of lipoprotein(a) in clinical laboratories worldwide. Towards this aim, the International Federation of Clinical Chemistry Working Group for the Standardization of Lipoprotein(a) Assays has initiated a standardization project involving a total of 33 diagnostic company and clinical chemistry laboratories from 12 countries. In Phase 1, the analytical performance of 40 lipoprotein(a) assay systems was evaluated by testing sera and manufactured lipoprotein(a) calibrator materials for precision, linearity, and parallelism. Twenty test systems were nonoptimized according to the results for a pooled serum, which tested nonlinear in 16 systems and imprecise in 4. Acceptable analytical properties and harmonization of lipoprotein(a) values were shown by some commercial calibrators, suggesting their possible use as reference materials. This study highlights the problems that currently occur for lipoprotein(a) measurement in existing assay systems.

Cite

CITATION STYLE

APA

Tate, J. R., Rifai, N., Berg, K., Couderc, R., Dati, F., Kostner, G. M., … Steinmetz, A. (1998). International federation of clinical chemistry standardization project for the measurement of lipoprotein(a). Phase I. Evaluation of the analytical performance of lipoprotein(a) assay systems and commercial calibrators. Clinical Chemistry, 44(8 I), 1629–1640. https://doi.org/10.1093/clinchem/44.8.1629

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free