A joint calibration technique for improving measurement accuracy of voltage and current probes during synchronous operation for RF-based plasma devices

11Citations
Citations of this article
8Readers
Mendeley users who have this article in their library.
Get full text

Abstract

This paper presents a joint calibration scheme for voltage (V) and current (I) probes that helps accurately resolve the voltage-current phase differences even when the difference is very close to 90°. The latter has been a major issue with V-I probes when used with miniature RF plasma devices such as the atmospheric pressure plasma jet (APPJ). Since the impedance of such miniature devices is predominantly capacitive, the phase difference between the voltage and current signals is very nearly 90°. It turns out, however, that when V-I probes are used with such devices without joint calibration, these frequently yield phase shifts over 90°. Also, since the power absorption is proportional to the resistive part of the impedance, it becomes very sensitive to the phase difference when it is close to ≈90°. Thus, it is important to be able to accurately resolve the phases. Post-calibration, V-I probes would be indispensable for the electrical characterization of APPJs for determining the average RF power Pav, plasma impedance Zp, etc. Typical post-calibration V-I data yield Zp ≈ 93.6 - j 1139 ω (81.5 - j 1173 ω) at Pav ≈9.8W (≈7.7W) for helium (argon) gas.

Cite

CITATION STYLE

APA

Mahreen, Ganguli, A., Gajula, V. P., Kar, S., & Sahu, D. (2022). A joint calibration technique for improving measurement accuracy of voltage and current probes during synchronous operation for RF-based plasma devices. Review of Scientific Instruments, 93(12). https://doi.org/10.1063/5.0124816

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free