This study performed a preliminary assessment on manufacturer's point calibration method by comparing it with a modified calibration setup by using an Air Bladder. Rather than having a subject participated in the calibration process, the modified calibration setup used the calibration bladder to mimic the force applied by the subject to generate a calibration fit. This assessment also includes a sensor performance evaluation by using both methods to monitor the overall performance of the sensor after every experiment. Result shows a large difference between manufacturer's calibration and modified calibration where the manufacturer's calibration tends to show an irregular sensor performance. © 2008 Springer-Verlag.
CITATION STYLE
Hamzah, H., Abu Osman, N. A., & Hasnan, N. (2008). Comparing manufacturer’s point calibration and modified calibration setup for F-scan insole sensor system: A preliminary assessment. In IFMBE Proceedings (Vol. 21 IFMBE, pp. 424–427). Springer Verlag. https://doi.org/10.1007/978-3-540-69139-6_108