Quality assurance and application of a 4D in vivo dosimetry system using a deformable lung phantom

0Citations
Citations of this article
7Readers
Mendeley users who have this article in their library.
Get full text

Abstract

A novel 4D in-vivo dosimetry detector (RADPOS) in conjunction with a deformable lung phantom has been used as a quality assurance tool for 4D radiotherapy. A treatment plan was created and delivered with the phantom in two breathing states and the RADOS system was used to verify dose and displacement of the tumour and lung volumes. The RADPOS measured the movement within a maximum deviation of 1.5 mm, and the dose within 4.2% of the total dose. The latest quality assurance and characterization tests of the RADPOS 4-D in-vivo dosimetry system are also presented in this work. © 2009 Springer-Verlag.

Cite

CITATION STYLE

APA

Cherpak, A., Serban, M., Seuntjens, J., & Cygler, J. E. (2009). Quality assurance and application of a 4D in vivo dosimetry system using a deformable lung phantom. In IFMBE Proceedings (Vol. 25, pp. 349–351). Springer Verlag. https://doi.org/10.1007/978-3-642-03474-9_98

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free