Noise, Sensitivity, and Resolution of Flow Cytometers

62Citations
Citations of this article
45Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

The sensitivity and resolution of flow cytometers are functions of the signal produced by a given particle as well as by the noise in the presence of which the signal is detected. The noise is primarily due to the fact that emission of light as well as its detection by photoelectric devises are stochastic processes. This fact leads to equations describing how resolution and sensitivity are limited by the magnitude of the signal, the background, and the photoelectron quantum yield of the detector. The equations are pointing to a method by which the signal and noise of a flow cytometer can be measured in absolute terms, as well as a way to determine fluorescence sensitivity without having to extrapolate to the noise level. The equations appear to be validated when applied to measuring data obtained with two different flow cytometers. © 1992 Wiley‐Liss, Inc. Copyright © 1992 Wiley‐Liss, Inc.

Cite

CITATION STYLE

APA

Steen, H. B. (1992). Noise, Sensitivity, and Resolution of Flow Cytometers. Cytometry, 13(8), 822–830. https://doi.org/10.1002/cyto.990130804