Numerical Simulations of Realistic Lead Trajectories and an Experimental Verification Support the Efficacy of Parallel Radiofrequency Transmission to Reduce Heating of Deep Brain Stimulation Implants during MRI

53Citations
Citations of this article
43Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Patients with deep brain stimulation (DBS) implants may be subject to heating during MRI due to interaction with excitatory radiofrequency (RF) fields. Parallel RF transmit (pTx) has been proposed to minimize such RF-induced heating in preliminary proof-of-concept studies. The present work evaluates the efficacy of pTx technique on realistic lead trajectories obtained from nine DBS patients. Electromagnetic simulations were performed using 4- and 8-element pTx coils compared with a standard birdcage coil excitation using patient models and lead trajectories obtained by segmentation of computed tomography data. Numerical optimization was performed to minimize local specific absorption rate (SAR) surrounding the implant tip while maintaining spatial homogeneity of the transmitted RF magnetic field (B 1+ ), by varying the input amplitude and phase for each coil element. Local SAR was significantly reduced at the lead tip with both 4-element and 8-element pTx (median decrease of 94% and 97%, respectively), whereas the median coefficient of spatial variation of B 1+ inhomogeneity was moderately increased (30% for 4-element pTx and 20% for 8-element pTx) compared to that of the birdcage coil (17%). Furthermore, the efficacy of optimized 4-element pTx was verified experimentally by imaging a head phantom that included a wire implanted to approximate the worst-case lead trajectory for localized heating, based on the simulations. Negligible temperature elevation was observed at the lead tip, with reasonable image uniformity in the surrounding region. From this experiment and the simulations based on nine DBS patient models, optimized pTx provides a robust approach to minimizing local SAR with respect to lead trajectory.

References Powered by Scopus

Analysis of tissue and arterial blood temperatures in the resting human forearm

4952Citations
N/AReaders
Get full text

Transmit SENSE

648Citations
N/AReaders
Get full text

Parallel Excitation with an Array of Transmit Coils

524Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Virtual clinical trials in medical imaging: A review

149Citations
N/AReaders
Get full text

MRI-Related Heating of Implants and Devices: A Review

96Citations
N/AReaders
Get full text

Reconfigurable MRI technology for low-SAR imaging of deep brain stimulation at 3T: Application in bilateral leads, fully-implanted systems, and surgically modified lead trajectories

52Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

McElcheran, C. E., Golestanirad, L., Iacono, M. I., Wei, P. S., Yang, B., Anderson, K. J. T., … Graham, S. J. (2019). Numerical Simulations of Realistic Lead Trajectories and an Experimental Verification Support the Efficacy of Parallel Radiofrequency Transmission to Reduce Heating of Deep Brain Stimulation Implants during MRI. Scientific Reports, 9(1). https://doi.org/10.1038/s41598-018-38099-w

Readers over time

‘19‘20‘21‘22‘23‘2405101520

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 12

50%

Researcher 7

29%

Professor / Associate Prof. 3

13%

Lecturer / Post doc 2

8%

Readers' Discipline

Tooltip

Engineering 13

50%

Medicine and Dentistry 5

19%

Neuroscience 4

15%

Physics and Astronomy 4

15%

Save time finding and organizing research with Mendeley

Sign up for free
0